MAINTENANCE & SUPPLIES, As Needed, RBC Transducer Cleaning Sequence
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-35
(5) When the detector chamber becomes empty, the message "Place 1 mL of
CELLCLEAN in the RBC Transducer and press Start Switch." will appear.
Sysmex
Ready
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Maintenance
<Clean RBC Transducer>
Place 1 mL of CELLCLEAN in the
RBC Transducer
and press Start Switch.
Return
Figure 5-7-10: Cleaning Transducer Message Window (2)
(6) Open the front cover and set the stop bar in position.
WARNING! When working with the front cover open, make sure that the stop
bar is in place.
(7) Remove the RBC transducer cap. See Figure 5-7-11.
Figure 5-7-11: Remove Transducer Cap
WARNING! CELLCLEAN (or 5% filtered sodium hypochlorite) is a strong
alkaline solution. Avoid contact with it. If you get it on your skin
or clothing, immediately wash it off with excess water.
Sysmex SF-3000 Operator's Manual -- Revised February 1996