8. ANALYSIS PARAMETERS
The SF-3000 uses 4 detection methods and 5 types of reagents to analyze the following
23 parameters:
Table 1-3: Analysis Parameters
Parameter Acronym Detection Method
White Blood Cell Count WBC Flow Cytometry by Semiconductor Laser
Red Blood Cell Count RBC DC
Hemoglobin Concentration HGB SLS hemoglobin
Hematocrit (true relative percentage volume
of erythrocytes)
HCT RBC Cumulative Pulse Height Detection
Mean Corpuscular (erythrocyte) Volume MCV Computed from RBC and HCT
Mean Corpuscular Hemoglobin MCH Computed from RBC and HGB
Mean Corpuscular Hemoglobin
Concentration
MCHC Computed from HCT and HGB
Platelet Count PLT DC
Neutrophil Percent NEUT%
Lymphocyte Percent LYMPH%
Monocyte Percent MONO% Flow Cytometry by Semiconductor Laser
Eosinophil Percent EO%
Basophil Percent BASO%
Neutrophil Count NEUT#
Lymphocyte Count LYMPH#
Monocyte Count MONO# Flow Cytometry by Semiconductor Laser
Eosinophil Count EO#
Basophil Count BASO#
RBC Distribution Width RDW-SD Computed
RBC Distribution Width RDW-CV Computed
Platelet Distribution Width PDW Computed
Mean Platelet Volume MPV Computed
Platelet Large Cell Ratio P-LCR Computed
NOTE:
Customers in the United States:
Platelet Distribution Width (PDW) and
Platelet Large Cell Ratio (P-LCR) parameters have not been approved
for reporting by the Food & Drug Administration (FDA). The obtained
values should be excluded from all print forms until reporting approval has
been obtained. See
Chapter 11
for information on excluding these
values.