MAINTENANCE & SUPPLIES, Replacements, Piercer (Auto Sampler)
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-55
2 . Installation
(1) Insert the new piercer, and retighten the thumbscrews (see Figure 5-8-18).
(2) Tighten the fixing screw until the piercer needle becomes free
(see Figure 5-8-19).
Tighten Two
Thumbscrews
Figure 5-8-18: Figure 5-8-19:
Tighten the Thumbscrews Tighten the Fixing Screw
WARNING! After tightening the needle fixing
screw, the piercer needle may be
exposed. Piercer Needle
(3) Reconnect the three tubes as shown in Figure 5-8-20.
Connect
Connect
Figure 5-8-20: Reconnect the Tubes