MAINTENANCE & SUPPLIES, Replacements, Piercer (Auto Sampler)
5-56 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(4) Return the piercing unit to its original position. The first step is to pull the unit
toward you to release the lock (Arrow in Figure 5-8-21). Next, lift the unit and
push it down to its original position (Arrows and ).
Piercing Unit
3
Figure 5-8-21: Set the Piercing Unit
CAUTION: Take care not to catch, pinch or bend the tubing when replacing the
piercing unit.
(5) Remove the holding metal securing the mixing mechanism from the cylinder rod
(Arrow in Figure 5-8-22). Then pull down the mixing mechanism (Arrow ).
(6) Tighten the two piercing unit thumbscrews to secure the piercing unit (see Figure
5-8-23).
Mixer Mechanism
Holding Metal
Up-Down
Air Cylinder
Tighten Two
Thumbscrews
Tighten
Tighten
Figure 5-8-22: Remove the Holding Metal Figure 5-8-23: Tighten the Thumbscrews