QUALITY CONTROL, Setting X M Control
Sysmex SF-3000 Operator's Manual -- Revised September 1995 6-11
(4) Pressing [Batch] keypad displays the numeric keypads to specify the number of
samples in a batch.
Use the numeric keys to enter the sample number and press [ENTER]. Pressing
[CANCEL] cancels the sample number setting without changing the number.
UL
LL
UL
LL
UL
LL
UL
LL
UL
LL
UL
LL
Li mit Data
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready Manual Mode Next No.1234567890123
Sampler
Q C
< XM Chart >
RBC
HGB
HCT
MCV
MCH
MCHC
XM
Sample No.
20
CANCEL
ENTER
7 8 9
4 5 6
1 2 3
0
C
ReturnFile Output Mean/SD
Target
/Limit
XM Start
/Stop
Delete More
XM Control
20Batch
Cancel Start
Figure 6-8: Sample No. Settings Window
(5) Press [Start] key in the X M Settings window to start X M Control. Pressing
[Cancel] cancels the settings and the setting window will close.
When the X M Control is started, " X M" will be displayed in the System Status
area in the top line of the LCD.
(6) To stop the X M Control, press [ X M Start/Stop] keypad. The X M Control
window as shown below appears.
XM Control
Cont. Stop
Figure 6-9: XM Control Window