QUALITY CONTROL, Sampler QC Analysis
Sysmex SF-3000 Operator's Manual -- Revised December 1995 6-45
X
M
QC02
Last
Status
Work
List
QC
Stored
Data
GP
Print
NO.
R B C P L T
D I F F W B C / B A S O
[g/dL]
[fL]
WBC
NEUT
LYMPH
MONO
EO
BASO
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
P-LCR
250fL 40fL
RBC Flag
PLT Flag
WBC Flag
< L-J Limit Error >
Sysmex
Ready
Sampler
Return
3
[x10 /uL]
[%]
[%]
[%]
[%]
[%]
6
[x10 /uL]
[g/dL]
[%]
[%]
[%]
[fL]
[fL]
[fL]
3
[x10 /uL]
[pg]
Manual Mode
Next No.1234567890123
HC GP DP SU
7.23
4.62
1.55
0.71
0.18
0.17
4.74
14.1
41.6
87.8
29.7
33.9
46.4
15.7
250
11.2
15.0
29.5
63.9
21.4
9.8
2.5
2.4
R:0001-01 15:48 1996/6/30
+
+
+
+
DP
Print
Replace
Reagent
Figure 6-39: Control Data Display (Auto Mode)
If the ID reader is installed, the lot number is read from the bar code label and the
data of the control sample is automatically stored in the specified QC files.
If the control lot number is registered for a QC file and the lot number is entered in
the Work List, the control vial is identified during the auto mode analysis and the
results will automatically be stored in the QC file. Refer to Chapter 3 to register the
control information in Work List.
Error Messages relating to QC analysis:
• < X Limit Error> or <L-J Limit Error>
This message appears in the system status area (upper portion of the LCD), when
the analyzed data has exceeded the control limit.
When the Limit Error is accompanied by the "Check Control Chart" message,
check the control chart and review the control data. Reanalyze the sample, if
needed.
The "+" or "-" sign in yellow back-light will be added to the parameter which is
outside the control limits.
The results are plotted on the QC chart as control data points, and are stored in the
QC file.
When it is accompanied with "Reanalyze Control" message, the analyzed data is
more than 3 times the limit. Reanalyze the sample.
The "+" or "-" sign in red back-light will be added to the parameter which is
outside the control limits.
The results are not plotted, or not stored as control data.