EasyManuals Logo

SYSMEX SF-3000 User Manual

SYSMEX SF-3000
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #348 background imageLoading...
Page #348 background image
TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-27
Rack Shift Function Error
Phenomena A sample rack did not shift in the measurement line properly.
(The rack shift function sensor does not turn ON within preset time limit.)
The "Reset Rack" error message also appears.
After the sample being processed is completed, the system returns to Ready.
Probable Cause 1) Rack movement was incorrect or incomplete.
2) Blood, etc. has interfered with the rack shift function sensor.
3) Actuator bearing of the rack position sensor is stuck.
Actuator bearing
4) The rack shift function sensor is defective.
Corrective Action 1) Check that nothing is obstructing rack movement, and remove any dirt or
foreign material from the rack moving area.
2) Apply lubricant to bearing.
3) Reset the rack in the right rack pool and press [Sampler], then [Start] key
to resume the analysis.
4) If the error persists, contact your Sysmex service representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX SF-3000 and is the answer not in the manual?

SYSMEX SF-3000 Specifications

General IconGeneral
BrandSYSMEX
ModelSF-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals