TROUBLESHOOTING
8-30 Sysmex SF-3000 Operator's Manual -- Revised September 1995
Rack Move Error 2
Phenomena A sample rack moved in the measurement line when the rack should not be
shifted.
The "Reset Rack" message also appears. When the sample being processed
is completed, the system returns to Ready.
Probable Cause 1) Rack was moved at incorrect timing.
2) Rack position sensor(s) are defective.
Corrective Action 1) The Reset the rack and press [Sampler], then [Start] key to resume the
analysis.
2) If the error still persists, contact your Sysmex service representative.
Rack Move Error 3
Phenomena A sample rack moved in the measurement line when STAT ready.
The "Reset Rack" message also appears. When the sample being processed
is completed, the system returns to Ready.
Probable Cause 1) Rack was moved when the instrument is ready for STAT analysis.
2) Rack position sensor(s) are defective.
Corrective Action 1) Reset the rack and press [Sampler], then [Start] key to resume the
analysis.
2) If the error still persists, contact your Sysmex service representative.