TROUBLESHOOTING
8-38 Sysmex SF-3000 Operator's Manual -- Revised September 1995
Laser Power Error
Phenomena The laser output level is outside of the control level.
An analysis error may occur, and the "Check Stored Data" message also
appears. Analysis cannot be continued until the error is resolved.
Probable Cause Laser unit is aged.
Corrective Action 1) Check the laser output operation using laser test program. (Refer to
Section 3.12: Test Laser Unit
in this chapter.)
2) Replacement of laser unit is necessary. Contact your Sysmex service
representative.
9 . Analysis Errors
< Data Error >
Phenomena Any parameter(s) of the analysis result exceeded the Sampler Stop Limit.
If the system is set to be stopped by Stop Limit Error, the system will be STAT
ready for the sample analysis in manual or capillary mode.
If the system is set to continue analysis when this error occurs, the system
will process the next sample.
Corrective Action 1) Check the Stored Data for the sample indicated with a flag "+" or "–" in red
back-light.
2) Resume the sampler operation by pressing [Sampler], then [Start] key on
the LCD.
< Analysis Error >
Phenomena An instrument abnormality was detected. A window containing analysis error
message(s) appears.
Corrective Action 1) Check the Stored Data for the sample indicated with "Func." or "Result" in
ERROR column in the stored data list display (P1).
2) Reset the rack. Press the [Sampler] keypad on the LCD and set manually
the Rack No. and the Tube Pos. No. for the next sample. Press
[Sampler], then [Start] key to resume the sampler operation.
Sysmex SF-3000 Operator's Manual -- Revised December 1995