TROUBLESHOOTING
8-44 Sysmex SF-3000 Operator's Manual -- Revised September 1995
10. Subprocessor Errors
Sheath Motor Controller Error
WB Asp Motor Controller Error
Rinse Motor Controller Error
Phenomena A communication problem exists between the CPU board and the motor(s).
An analysis error may occur, and analysis cannot be continued until the error
is resolved.
Probable Cause CPU operation faulty due to sudden noise interference.
Corrective Action 1) Move the source of external noise (such as radio, centrifuge, temperature
controlled water bath, etc.) well away from the instrument.
2) Turn the Main Unit power OFF, then ON again.
3) If the error persists, contact Sysmex service representative.
Sampler Com. Error
ID Unit Com. Error
Phenomena A communication problem exists between the CPU board and the sampler or
ID reader.
An analysis error may occur, and analysis cannot be continued until the error
is resolved.
Probable Cause 1) CPU operation faulty due to sudden noise interference.
2) The ID reader or sampler is disconnected or improperly connected.
Corrective Action 1) Move the source of external noise (such as radio, centrifuge, temperature
controlled water bath, etc.) well away from the instrument.
2) Verify the cable between the Main Unit and the Sampler/ID reader is
securely connected.
3) Turn the Main Unit power OFF, then ON again.
4) Check the ID reader function using test program. Refer to
Section 3.7:
Test Bar Code Reader
in this chapter.
5) If the error persists, contact Sysmex service representative.