INTRODUCTION
1-24 Sysmex SF-3000 Operator's Manual -- Revised March 2001
Table 1-4: System Specifications (Continued)
Carryover WBC 3.0% or less
RBC 1.5% or less
HGB 1.5% or less
HCT 1.5% or less
PLT 1.5% or less
Maximum Allowable Difference
between Manual Open Mode and
Auto Mode
WBC Within ± 0.4 x 10
3
/µL or ± 5%
RBC Within ± 0.02 x 10
6
/µL or ± 2%
HGB Within ± 0.2 g/dL or ±2%
HCT Within ± 0.3HCT% or ± 2%
PLT Within ± 20 x 10
3
/µL or ± 7%
NEUT% Within ± 5.0% NEUT%
LYMPH% Within ± 4.0% LYMPH%
MONO% Within ± 3.0% MONO%
EO% Within ± 2.0% EO%
BASO% Within ± 1.0% BASO%
Display Range WBC 0.00 - 999.99 x 10
3
/µL
RBC 0.00 - 99.99 x 10
6
/µL
HGB 0.0 - 30.0 g/dL
HCT 0.0 - 100%
PLT 0 - 9999 x 10
3
/µL
Peripheral Interface Serial port for Host Computer
Parallel port for Data Printer (DP-510, DP-490)
Parallel port for Graphic Printer (DeskJet 560C/560J, LaserJet,
Canon BJC 400J, LP-1500, LP-1600)
Dimensions H x W x D (mm), and
Weight
Main Unit Approx. 580 x 451 x 600, 58 kg
Pneumatic Unit Approx. 195 x 395 x 333, 15.5 kg
Sampler Unit Approx. 580 x 280 x 302, 14.5 kg
Acceptable Barcodes CODABAR, NW-7, CODE-11, CODE-39, CODE-128,
Interleaved 2 of 5, JAN
Classification Type of protection against electric shock Class I equipment
Degree of protection against electric shock Type B applied part
Heat Compensation Required Approx. 2047 BTU/h (516 kcal/h)