EasyManua.ls Logo

SYSMEX SF-3000 - Page 36

SYSMEX SF-3000
605 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INTRODUCTION
Sysmex SF-3000 Operator's Manual -- Revised September 1995 1-23
Table 1-4: System Specifications (Continued)
Differential Accuracy The instrument differential is compared by using 100 or more fresh samples
to the reference 400 cell manual differential (200 counts with 2 Lab. Techs.),
as per NCCLS H-20A. MONO% is compared by using the monoclonal
antibodies. Results are displayed as a range of values that are the mean
differences between the methods, expressed to a 95% confidence interval.
NEUT% Coefficient correlation r 0.90
LYMPH% Coefficient correlation r 0.90
MONO% Coefficient correlation r 0.75
EO% Coefficient correlation r 0.80
BASO% Coefficient correlation r 0.50
Additionally, the mean difference from that obtained on the standard
instrument should be the following:
NEUT% within ± 3.0 NEUT%
LYMPH% within ± 3.0 LYMPH%
MONO% within ± 2.0 MONO%
EO% within ± 1.0 EO%
BASO% within ± 1.0 BASO%
Linearity
for Closed and
Open mode
WBC within ± 0.30 x 10
3
/µL or within 3%
for range 1.00-99.99 x 10
3
/µL
RBC within ± 0.03 x 10
6
/µL or within 3%
for range 0.30-9.99 x 10
6
/µL
HGB within ± 0.2 g/dL or within 2%
for range 0.1-25.0 g/dL
HCT within ± 1.0 HCT% or within 3%
for range 10.0-60.0 HCT%
PLT within ± 10 x 10
3
/µL or within 5%
for range 10-999 x 10
3
/µL
(Condition: RBC count is less than approx. 7.00 x 10
6
/µL)
Linearity
for Capillary mode
WBC within ± 0.50 x 10
3
/µL or within 5%
for range 1.00-99.99 x 10
3
/µL
RBC within ± 0.06 x 10
6
/µL or within 6%
for range 1.00-9.99 x 10
6
/µL
HGB within ± 0.7 g/dL or within 7%
for range 2.0-25.0 g/dL
HCT within ± 2.0 HCT% or within 6%
for range 10.0-60.0 HCT%
PLT within ± 20 x 10
3
/µL or within 10%
for range 10-999 x 10
3
/µL
(Condition: RBC count is less than approx. 7.00 x 10
6
/µL)
NOTE: The Linearity range is different from the Display range. Consult the
Linearity specification for the reportable range for each parameter. The
Display range is given on the next page.

Table of Contents

Related product manuals