1-22 Sysmex SF-3000 Operator's Manual -- Revised September 1995
Table 1-4: System Specifications (Continued)
Precision
for Closed and
Open mode
Precision Characteristics are determined during instrument evaluation using
one fresh normal whole blood sample analyzed ten times consecutively in
Closed Sampling Auto Mode or Open Sampling Manual Mode.
WBC CV 3.0% (when WBC 4.0 x 10
3
/µL)
NEUT% CV 8.0% (when NEUT% 30.0% and WBC 4.0 x 10
3
/µL)
LYMPH% CV 8.0% (when LYMPH% 15.0% and WBC 4.0 x 10
3
/µL)
MONO% CV 20.0% (when MONO% 5.0% and WBC 4.0 x 10
3
/µL)
EO% CV 25.0% or within ± 1.5 EO% (when WBC 4.0 x 10
3
/µL)
BASO% CV 40.0% or within ± 1.0 BASO% (when WBC 4.0 x 10
3
/µL)
NEUT# CV 8.0% (when NEUT# 1.20 x 10
3
/µL)
LYMPH# CV 8.0% (when LYMPH# 0.60 x 10
3
/µL)
MONO# CV 20.0% (when MONO# 0.20 x 10
3
/µL)
EO# CV 25.0% or within ± 0.12 x 10
3
/µL
BASO# CV 40.0% or within ± 0.06 x 10
3
/µL
RBC CV 1.5% (when RBC 4.00 x 10
6
/µL)
HGB CV 1.5%
HCT CV 1.5%
MCV CV 1.5%
MCH CV 1.5%
MCHC CV 2.0%
PLT CV 5.0% (when PLT 100 x 10
3
/µL)
RDW-CV CV 3.0%
RDW-SD CV 3.0%
PDW CV 10.0%
MPV CV 4.0%
P-LCR CV 18.0%
Precision
for Capillary mode
Precision Characteristics are determined during instrument evaluation using
one fresh normal capillary blood sample analyzed ten times
consecutively in Open Sampling Manual Mode.
WBC CV 9.0% (when WBC 4.0 x 10
3
/µL)
RBC CV 4.5% (when RBC 4.00 x 10
6
/µL)
HGB CV 4.5%
HCT CV 4.5%
MCV CV 4.5%
MCH CV 4.5%
MCHC CV 6.0%
PLT CV 15.0% (when PLT 100 x 10
3
/µL)
Accuracy
for Closed mode
for Open mode
WBC within ± 3% or within 0.20 x 10
3
/µL
RBC within ± 2% or within 0.03 x 10
6
/µL
PLT within ± 5% or within 10 x 10
3
/µL
Accuracy
for Capillary mode
WBC within ± 10%
RBC within ± 8%
PLT within ± 12%