TROUBLESHOOTING
8-54 Sysmex SF-3000 Operator's Manual -- Revised September 1995
Paper Empty
Phenomena The paper for the printer has run out.
Output to the printer is impossible until the error is resolved.
Probable Cause The printer is out of paper.
Corrective Action Supply the printer paper.
Power/Cable OFF
Phenomena The printer is powered OFF or the interface cable is disconnected.
The print out to GP or DP cannot be performed.
Probable Cause 1) The printer is not powered ON.
2) The interface cable is disconnected.
Corrective Action 1) Power ON the printer.
2) Verify that the interface cable is securely connected.
3) If the error persists, contact Sysmex service representative.