TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-53
Not Assigned
Phenomena The output to GP or DP is assigned when the printer is not specified for
connection.
Output to GP or DP is impossible until the error is resolved.
Probable Cause GP or DP has not been specified in Settings program.
Corrective Action 1) Set the GP or DP to "Connected". Refer to
Chapter 11, Section 5.2: GP
Setting
or
Section 5.3: DP Setting
for the procedures.
2) If the error persists, contact Sysmex service representative.
Off Line/Device Error
Phenomena The output to GP or DP is assigned when the printer is in the off line condition
or when a printer error occurs.
Output to GP or DP is impossible until the error is resolved.
Probable Cause 1) GP or DP printer power is OFF or is off-line.
2) A printer hardware error occurred.
Corrective Action 1) Check the printer power is ON.
2) Set the printer in the on line condition.
3) If the error persists, contact Sysmex service representative.
NOTE: "GP Off Line" error message will disappear automatically when the
printer becomes on-line condition. However the beep sound will not
stop until [Alarm Reset] key is pressed.