TROUBLESHOOTING
8-58 Sysmex SF-3000 Operator's Manual -- Revised September 1995
Clean the WBC Detector
Probable Cause WBC Detector counter has reached 600 or 1 week passed after the previous
maintenance of WBC Detector.
Corrective Action 1) Clean the WBC Detector. Refer to
Chapter 5, Section 4.2: Clean WBC
Detector
.
2) Reset the WBC Detector counter. Refer to
Chapter 5, Section 9: Reset
the Counter.
Replace Piercer
Probable Cause Piercer counter has reached 30 000.
Corrective Action 1) Replace the piercer. Refer to
Chapter 5, Section 8.2: Replace the Piercer
.
2) Reset the Piercer counter. Refer to
Chapter 5, Section 9: Reset the
Counter.
Sysmex SF-3000 Operator's Manual -- Revised July 1996