TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-57
16. Maintenance Instruction Messages
When the counters for each unit reach the cycle count needing cleaning or replacement,
an instruction message will be displayed at the system power ON .
Clean the Orifice
Probable Cause Vacuum Orifice counter has reached 2500 or 1 month passed after the
previous maintenance of Orifice.
Corrective Action 1) Clean the Orifice. Refer to
Chapter 5, Section 5.1: Clean Orifice
.
2) Reset the Vacuum Orifice counter. Refer to
Chapter 5, Section 9: Reset
the Counter.
Clean the SRV
Probable Cause SRV counter has reached 2500 or 1 month passed after the previous
maintenance of SRV.
Corrective Action 1) Clean the SRV. Refer to
Chapter 5, Section 5.3: Clean SRV
.
2) Reset the SRV counter. Refer to
Chapter 5, Section 9: Reset the Counter.
Clean the Waste Chamber
Probable Cause Waste Chamber counter has reached 2500 or 1 month passed after the
previous maintenance of Waste Chamber.
Corrective Action 1) Clean the Waste Chamber. Refer to
Chapter 5, Section 5.2: Execute
Clean Waste Chamber Sequence
.
2) Reset the Waste Chamber counter. Refer to
Chapter 5, Section 9: Reset
the Counter.