TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-93
P3. Sensor Status
Use this program to check pressure and vacuum. Refer to Chapter 9: Adjustment for the
procedures.
Also the status of the solenoid valves and sensors, and the temperature in each unit can
be observed using this screen.
The SV (solenoid valve) number displayed in yellow back-light is in the ON status.
The sensor number displayed in green back-light is in the ON status.
The SV/Sensor status is updated every second.
STATUS
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Next No.1234567890123
Sampler
Manual Mode
Return
< SV/Sensor Status >
2
12
22
32
42
52
62
72
82
1
11
21
31
41
51
61
71
81
3
13
23
33
43
53
63
73
83
4
14
24
34
44
54
64
74
84
5
15
25
35
45
55
65
75
85
6
16
26
36
46
56
66
76
86
7
17
27
37
47
57
67
77
87
8
18
28
38
48
58
68
78
88
9
19
29
39
49
59
69
79
89
10
20
30
40
50
60
70
80
90
S V
1
11
21
31
2
12
22
32
3
13
23
33
4
14
24
34
5
15
25
35
6
16
26
36
7
17
27
37
8
18
28
38
9
19
29
39
10
20
30
40
S E N S O R
Press
Press
Press
Vacuum
2.0k :
0.6k :
0.5k :
250 :
2.000kg/cm2
0.605kg/cm2
0.510kg/cm2
250 mmHg
RBC Detector
WBC Detector
React Chamber
Sheath Reagent
: 27.0 °C
: 30.0 °C
: 35.0 °C
: 30.0 °C
HGB : 12.00
←→
Figure 8-36: Status Display - SV/Sensor Status Screen