TROUBLESHOOTING
8-92 Sysmex SF-3000 Operator's Manual -- Revised September 1995
P2. Analysis Process
Sequential or overlap analysis of up to three samples can be carried out within the
instrument. Analysis status of each sample is displayed on the "Analysis Process"
screen. Normally, the status of the last 4 samples is displayed. The sample displayed in
the left column is a newly analyzed sample and the status is updated every second.
The sample is identified by sequence number, ID No. and rack position number on the
top three lines of the matrix. A hyphen will appear in "RACK" if the sample is analyzed
in Manual mode. The display has 10 processing steps.
The symbols used to report the status are shown below with their meaning:
: ENTRY
Step is expected.
: RUN
Step is in progress.
: FINISH NORMAL
Completed without error.
: INTERRUPTION
Step is interrupted.
: FINISH ABNORMAL
Completed with error.
: (blank) The process has not begun.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Status
< Analysis Process >
Sequence
ID No.
Rack
Aspirate
Dilute
RBC Count
WBC/BASO Count
DIFF Count
RBC Analyze
WBC/BASO Analyze
DIFF Analyze
Edit
Store
13
123-567-90123
0000-00
12
123-567-90122
0000-00
11
123-567-90121
0000-00
10
123-567-90120
0000-00
: Entry
: Interruption
: Run
: Finish Abnormal
: Finish Normal
Return
Figure 8-35: Status Display - Analysis Process Screen