SYSTEM SETUP, DP Setting
Sysmex SF-3000 Operator's Manual -- Revised September 1995 11-39
Status
DP-510 DP-490
Model
4
Decimal Point
Date
Print Type
/ Space None
Delimiter
of Date
Not Printed
< DP Settings >
More
Not Connected
Sample No. Digits
Printed
Connected
6 8 10 13
yymmdd mmddyy ddmmyy ddmmmmdd
Quit
Return
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
HC
Setting
GP
Setting
DP
Setting
√
√
Figure 11-31: DP Setting Screen - 1
Status : When the Data Printer is connected, select "Connected". If
"Not Connected" is selected, "DP" will not appear in the
LCD system status display area. (The factory setting is
"Connected".)
Model : Select the model "DP-510" or "DP-490". (The factory
setting is "DP-510".)
Sample No. Digits : Set the maximum number of digits for sample ID Nos. ID
Nos. can be set 4, 6, 8, 10 or 13 digits. ID Nos. will be
printed in order of the lowest number of digits to the highest
number of digits with the maximum number of digits being
the value set here. (The factory setting is "13 digits".)
Decimal Point : Specify whether the decimal point is to be printed in the data
or not. Select "Printed" or "Not Printed". (The factory
setting is "Printed".)
Date Print Type : Select one of the date print formats. (The factory setting is
"mmddyy".)
Delimiter of Date : Three different delimiters can be used between the figures for
day, month and year. (The factory setting is "/".)