INSTALLATION, Power ON and System Setup
B-34 Sysmex SF-3000 Operator's Manual -- Revised September 1995
Mode-to-Mode Calibration
(1) When a CP Sampler Unit is installed, analyze 10 normal fresh patient samples both
in the Open Manual mode and the Closed Sampler mode.
(2) Calculate the mean of those samples in the Open Manual mode and the mean of
those samples in the Closed Sampler mode. Verify that either the percent
difference or the absolute difference value is within the following specification.
Specification:
WBC: within ±5%, or within ±0.40 x 10
3
/µL
RBC: within ±2%, or within ±0.02 x 10
6
/µL
HGB: within ±2%, or within ±0.2 g/dL
HCT: within ±2%, or within ±0.3 Hct%
PLT: within ±7%, or within ±20 x 10
3
/µL
NEUT%: within ±5.0 NEUT%
LYMPH%: within ±4.0 LYMPH%
MONO%: within ±3.0 MONO%
EO%: within ±2.0 EO%
BASO%: within ±1.0 BASO%
(3) If not found within the specification, obtain and set the calibration value.
ID Reader
NOTE: When adjusting the ID Bar-Code Reader, contact your Sysmex
service representative for his/her instruction and assistance.
(1) Pull out the Sampler Unit by approx. 10 cm, and remove the Main Unit Front
Panel-B Assy.
CAUTION: DO NOT open the front panel or try to remove the Sampler Unit
without instruction and assistance of Sysmex service representative.
DO NOT pull the Sampler Unit more than 10 cm to avoid tubing
disconnection.
(2) Push and refit the Sampler Unit.
CAUTION: When refit the Sampler Unit, take care not to pinch the tubing.