SAMPLE PROCESSING, Auto Mode Procedures
2-36 Sysmex SF-3000 Operator's Manual -- Revised September 1995
CAUTION: If all the following conditions are met, the sample is assumed blank and
the analysis result will be neither stored nor judged against any
flagging criteria.
WBC < 1.00 x 10
3
/µL
RBC < 0.30 x 10
6
/µL
HGB < 1.0 g/dL
PLT < 20 x 10
3
µL
5 . Output Analysis Results
If automatic output has been set, the analysis results will be sent to the Graphic Printer,
Data Printer, and/or Host Computer.
If automatic output has not been set, the operator must select the data and destination for
the output. See Chapter 4: Result Interpretation/Reports.
6 . Perform Post-Analysis Procedures
Residual blood and reagent can cause detector aperture to clog and protein to build up in
aspiration lines. If the instrument is not to be used for a while, press the [Auto Rinse]
key in the root menu to execute a background check.
NOTE: If the [Auto Rinse] key is not displayed on the LCD, press [More] key
to change the root menu display.