SAMPLE PROCESSING, Auto Mode Procedures
Sysmex SF-3000 Operator's Manual -- Revised September 1995 2-35
• Stat Analysis in Auto Mode
(1) Press the [Sampler Stop] key to interrupt the sampler operation. When the
messages "Ready" and "STAT" appear on the LCD, manual analysis can be
performed.
(2) Press the Next No. keypad in the top line of the LCD. The Sample No. Setting
screen will appear.
(3) The selected analysis mode is backlit in light-blue. Press [Manual Mode] to select
Manual mode, if needed.
(4) Input the sample ID number for the Stat analysis with the numerical keys, and
press [ENTER]. The sample No. is set and the Sample No. Setting screen will
disappear.
(5) Process the sample as if it were a routine manual mode sample. While the sample
is being processed, the message "Running" will appear on the LCD.
(6) After the Stat analysis is completed, the message "Ready" will appear on the LCD
again; press the [Sampler] key.
(7) The Sampler Start Check window will appear. Press [Start] key to resume the
sampler operation.
4 . Display Analysis Results
After analysis is completed (approximately 85 seconds after the sample is aspirated), the
results are displayed on the screen.
XM
12345
Last
NO.
R B C P L T
D I F F W B C / B A S O
[g/dL]
[fL]
WBC
NEUT
LYMPH
MONO
EO
BASO
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
P-LCR
250fL 40fL
RBC Flag
PLT Flag
WBC Flag
DP:1234567890123
Sysmex
Ready
Sampler
3
[x10 /uL]
[%]
[%]
[%]
[%]
[%]
6
[x10 /uL]
[g/dL]
[%]
[%]
[%]
[fL]
[fL]
[fL]
3
[x10 /uL]
[pg]
Manual Mode
Next No.1234567890123
HC GP D P SU
7.23
4.62
1.55
0.71
0.18
0.17
4.74
14.1
41.6
87.8
29.7
33.9
46.4
15.7
213
9.8
10.3
24.3
63.9
21.4
9.8
2.5
2.4
R:1000-10
15:48 1996/6/30
Status
Work
List
QC
Stored
Data
GP
Print
Return
DP
Print
Replace
Reagent
Figure 2-27: Analysis Results Screen (Auto Mode)