Remote control commands
R&S
®
RTP
1141User Manual 1337.9952.02 ─ 12
SEARch:SOURce 'RFFESearch1',SBUS1
SEARch:TRIGger:RFFE:TYPE 'RFFESearch1',STOP
SEARch:TRIGger:RFFE:SADD:COND 'RFFESearch1', EQU
SEARch:TRIGger:RFFE:SADD:MIN 'RFFESearch1', 'XXX1'
RUNSingle;*OPC? //asynchronous command
SEARch:ALL 'RFFESearch1';*OPC?
SEARch:RESult:RFFE:FCOunt? 'RFFESearchCh1'
SEARch:RESult:RFFE:SEQuence1:STATe? 'RFFESearchCh1'
<--OK //First sequence is valid
SEARch:RESult:RFFE:SEQuence1:TYPE? 'RFFESearchCh1'
<--ERRL //Extended Register Read Long sequence
22.4 Frequently used parameters and suffixes
This chapter describes in general those parameters and suffixes that are used in sev-
eral subsystems.
22.4.1 Waveform suffix
The numeric waveform suffix is used in some commands, for example, to indicate the
source waveform number from which the reference level is taken, and to assign color
tables to waveforms.
Depending on the command, not all suffix values are supported. For example, in
REFLevel commands, only suffixes 2 to 21 are allowed. The range of supported suffix
numbers is indicated in the description of the individual commands.
Suffix 1
Suffix 1 means that no waveform is assigned. The first waveform C1W1 corresponds
to suffix number 2.
Waveform number Description
1 None
2 C1W1 (channel 1, waveform 1)
3 C1W2 (channel 1, waveform 2)
4 C1W3 (channel 1, waveform 3)
5 C2W1 (channel 2, waveform 1)
6 C2W2 (channel 2, waveform 2)
Frequently used parameters and suffixes

Table of Contents