EasyManua.ls Logo

Rohde & Schwarz RTP Series - MDIO Trigger

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTP
793User Manual 1337.9952.02 ─ 12
2. Select the tab of the bus you want to set up, for example "SB1".
3. Select the "Setup" tab.
4. In the "Setup" tab, set the "Protocol" > "MDIO".
5. Select the waveform for the "Clock" and "Data" lines.
6. Set the logical thresholds.
7. Enable "Decode".
13.17.3 MDIO trigger
13.17.3.1 MDIO trigger settings
Access: [Protocol] > "Shortcuts" tab > "Setup trigger"
In this section, all trigger settings are described. The user interface of the instrument
displays only appropriate settings and guides you through the trigger setup.
For a list of supported trigger conditions, refer to the data sheet.
Make sure that:
In the [Protocol] > "Setup" tab:
The correct "Protocol" is selected.
The data sources of the serial bus are channel signals.
In the "Trigger" > "Setup" tab:
The "Trigger on" = "Single event"
The trigger "Source" = "Serial bus".
The correct "Serial bus" is selected.
MDIO (option R&S