Triggers
R&S
®
RTP
224User Manual 1337.9952.02 ─ 12
The timeout trigger checks if the signal stays above or below the threshold voltage for
a specified time lapse. In other words, the trigger occurs if the trigger source does not
have the expected transition within the specified time.
Trigger level
Sets the voltage level for the trigger condition. You can also drag the trigger level
marker on the display (TA or TB on the right edge of the display). The range of the trig-
ger level is limited in a way so that always a hysteresis for stable trigger conditions is
available.
Remote command:
TRIGger<m>:LEVel<n>[:VALue] on page 1255
Range
Selects the relation of the signal level to the trigger level:
"Stays high"
The signal level stays above the trigger level.
"Stays low"
The signal level stays below the trigger level.
"High or low"
The signal level stays above or below the trigger level.
Remote command:
TRIGger<m>:TIMeout:RANGe on page 1266
Time
Defines the time limit for the timeout at which the instrument triggers.
Remote command:
TRIGger<m>:TIMeout:TIME on page 1267
6.3.8 Interval
Access: "Menu">"Trigger" > "Setup" tab > "Type = Interval"
The interval trigger analyzes the time between two pulses.
The interval trigger can analyze either rising or falling edges, but searching for an inter-
val is also possible for both edges at the same time ("Either").
Trigger types