Remote control commands
R&S
®
RTP
1199User Manual 1337.9952.02 ─ 12
ACQuire:SEGMented:AUToreplay <ReplayAfterAcq>
If enabled, the instrument starts processing and displaying the data as soon as the
acquisition series is captured completely. Depending on the number of acquisitions, it
may take some time until the acquisition series is displayed. If the setting is disabled,
the instrument only captures the data and stores it in the sample memory.
Parameters:
<ReplayAfterAcq> ON | OFF
*RST: ON
Usage: Asynchronous command
22.8.5
Vertical
CHANnel<m>:STATe.................................................................................................... 1199
CHANnel<m>:COUPling............................................................................................... 1199
CHANnel<m>:SCALe................................................................................................... 1200
CHANnel<m>:RANGe.................................................................................................. 1200
CHANnel<m>:POSition.................................................................................................1201
CHANnel<m>:OFFSet..................................................................................................1201
CHANnel<m>:INVert.................................................................................................... 1202
CHANnel<m>:BANDwidth.............................................................................................1202
CHANnel<m>:CPLing...................................................................................................1203
CHANnel<m>:IMPedance.............................................................................................1203
CHANnel<m>:OVERload.............................................................................................. 1203
CHANnel<m>:STATe <State>
Switches the channel signal on or off.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
CHANnel<m>:COUPling <Value>
Selects the connection of the indicated channel signal.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Value> DC | DCLimit | AC