Remote control commands
R&S
®
RTP
1314User Manual 1337.9952.02 ─ 12
Suffix: 
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Polarity> POSitive | NEGative
*RST:  POSitive
Usage:  Asynchronous command
TRIGger<m>:OUT:PLENgth 
<PulseLength>
Sets the length of the trigger out pulse.
Suffix: 
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<PulseLength> Range:  4E-9  to  0.06
Increment:  20E-9
*RST:  100E-9
Default unit: s
Usage:  Asynchronous command
TRIGger<m>:OUT:DELay <Delay>
Sets the delay of the first pulse edge to the trigger point.
The setting is not available if a mask test or measurement is running and the on-viola-
tion event is set to trigger out pulse.
Suffix: 
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Delay> Range:  1.8E-6  to  512E-3
Increment:  1E-9
*RST:  1.8E-6
Default unit: s
Usage:  Asynchronous command
22.9.21  Actions on trigger
TRIGger<m>:EVENt:BEEP........................................................................................... 1315
TRIGger<m>:EVENt:PRINt........................................................................................... 1315
TRIGger<m>:EVENt:WFMSave.....................................................................................1315
TRIGger<m>:EVENt:RUNexec......................................................................................1315
TRIGger<m>:OUT:ACTion............................................................................................ 1316
Trigger