Remote control commands
R&S
®
RTP
1247User Manual 1337.9952.02 ─ 12
22.8.10 Digital filter
CHANnel<m>:DIGFilter:STATe...................................................................................... 1247
CHANnel<m>:DIGFilter:CUToff......................................................................................1247
CHANnel<m>:DIGFilter:STATe <State>
Enables the DSP filter for input channels.
Suffix:
<m>
.
1..4
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
CHANnel<m>:DIGFilter:CUToff
<CutOffLP>
Sets the limit frequency of the lowpass filter for input channels.
Suffix:
<m>
.
1..4
Selects the input channel. One filter is applied to a pair of chan-
nels - one filter for channels 1 and 2 and another filter for chan-
nels 3 and 4 (only 4-channel models).
Parameters:
<CutOffLP> Range: 100E+3 to instrument bandwidth
*RST: 1E+6
Default unit: Hz
Available filter values are:
100E+3, 200E+3, 300E+3, 500E+3, 1.0E+6, 2.0E+6, 3.0E+6,
5.0E+6, 10E+6, 20E+6, 30E+6, 50E+6, 100E+6, 500E+6, 1.0E
+9, 1.5E+9, 2.0E+9, 2.5E+9, 3.0E+9, 3.5E+9, 4.0E+9, 4.5E+9,
5.0E+9, 5.5E+9, 6.0E+9, 6.5E+9, 7.0E+9, 7.5E+9, 8.0E+9, 8.5E
+9, 9.0E+9, 9.5E+9, 10.0E+9, 10.5E+9, 11.0E+9, 11.5E+9,
12.0E+9, 12.5E+9, 13.0E+9, 13.5E+9, 14.0E+9, 14.5E+9, 15.0E
+9, 15.5E+9, 16.0E+9
Usage: Asynchronous command
22.8.11 Skew
CHANnel<m>:SKEW:MANual....................................................................................... 1248
CHANnel<m>:SKEW:TIME........................................................................................... 1248
PROBe<m>:SKEState..................................................................................................1248
CHANnel<m>:SKEW:EXTended....................................................................................1248
Acquisition and setup