EasyManua.ls Logo

Rohde & Schwarz RTP Series - 100 BASE-T1 Trigger

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTP
648User Manual 1337.9952.02 ─ 12
13.10.3 100BASE-T1 trigger
13.10.3.1 100BASE-T1 trigger settings
Access: [Protocol] > "Shortcuts" tab > "Setup trigger"
In this section, all trigger settings are described. Their availability on the instrument
depends on the selected trigger type. The user interface of the instrument displays only
appropriate settings and guides you through the trigger setup.
For a list of supported trigger conditions, refer to data sheet.
Make sure that:
In the [Protocol] > "Setup" tab:
The correct "Protocol" is selected.
The data sources of the serial bus are channel signals.
In the "Trigger" > "Setup" tab:
The "Trigger on" = "Single event"
The trigger "Source" = "Serial bus".
The correct "Serial bus" is selected.
Type
Selects the trigger type for 100BASE-T1 analysis.
Remote command:
TRIGger<m>:HBTO:TYPE on page 1755
Frame Start ← Type
Triggers on the start of frame.
MAC Frame ← Type
Triggers on a Media Access Control (MAC) frame. You can specify:
Ethernet 100BASE-T1 (option R&S
RTP-57)

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals