Waveform analysis
R&S
®
RTP
307User Manual 1337.9952.02 ─ 12
Threshold
Defines the minimum frequency magnitude for which phases are calculated. This set-
ting is only available if "Suppression" is enabled.
Remote command:
CALCulate:MATH<m>:FFT:PHASe:THReshold on page 1454
Vertical maximum
Defines the maximum value on y-axis for spectrum displays. Only available for "Man-
ual" scale mode.
Remote command:
CALCulate:MATH<m>:VERTical:MAXimum on page 1339
Vertical range
Defines the range of FFT values to be displayed.
Remote command:
CALCulate:MATH<m>:VERTical:RANGe on page 1339
7.3.7 Math arithmetic
Access: [Math] > "Arithmetic" tab.
In this tab you can specify the waveform arithmetic for the math waveforms.
Mode
Waveform arithmetic builds the resulting waveform from several consecutive acquisi-
tions and subsequent math calculations of the signal. For details, see "Wfm Arithmetic"
on page 152.
"Original"
The original results are displayed.
"Envelope"
The envelope curve of all acquired and calculated results is dis-
played.
"Average"
The average of all acquired and calculated results is displayed.
Mathematics