Remote control commands
R&S
®
RTP
1316User Manual 1337.9952.02 ─ 12
●
EXECutable:PARameter on page 1546
●
EXECutable:WDIRectory on page 1547
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<RunExecutable> NOACtion | TRIGger
*RST: NOACtion
Usage: Asynchronous command
TRIGger<m>:OUT:ACTion <StateAction>
Enables/disbales the trigger out signal that is provided to the [Trigger Out] connector
on the rear panel when a trigger occurs.
Suffix:
<m>
.
1..3
Parameters:
<StateAction> NOACtion | TRIGger
*RST: NOACtion
Usage: Asynchronous command
22.9.22 Zone trigger (option R&S RTP-K19)
TRIGger<m>:ZONE:EXPRession[:DEFine] <LogiExpr>
Defines the zone trigger.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<LogiExpr> String with the logical expression
Example: Prerequisite: mask tests MT1 and MT2 are defined, and also
Zone1.
TRIGger:ZONE:EXPRession 'MT1 and MT2 and Zone1'
TRIGger:ZONE:EXPRession?
<-- MT1 and MT2 and Zone1
Usage: Asynchronous command
Firmware/software: FW 3.20