Mask testing
R&S
®
RTP
435User Manual 1337.9952.02 ─ 12
Figure 10-5: Waveform mask with vertical width = 0.5 div and vertical position = -0.5 div
Remote command:
MTESt:WFMRescale:YPOSition on page 1471
10.2.3 Actions on event
The settings in this tab define what happens when the mask test has failed or when it
has passed successfully. Furthermore, you can reset all totals and results in the "Mask
Test" result boxes.
Most actions can be initiated either on failure or on success:
On violation
The action is initiated when the fail criteria is fulfilled.
On successful completion
The action is initiated when the [Single] acquisition has finished and the fail criteria
is not fulfilled - the fail condition and violation tolerance limit have not been
reached.
There are two usual test practices:
Testing a defined number of waveforms against the mask and initiate an action
when the acquisition cycle has been completed without failure:
Set the number of acquisitions to be tested: "Average count (N-single count)"
Start [Single]
Testing a continuous acquisition or a defined number of waveforms against the
mask and initiate an action when the fail criteria is fulfilled
Mask test settings