Remote control commands
R&S
®
RTP
1469User Manual 1337.9952.02 ─ 12
MTESt:SEGMent:RESCale:YOFFset <MaskTestName>, <MaskSegmIdx>,<OffsetY>
MTESt:SEGMent:RESCale:YOFFset? <MaskTestName>, <MaskSegmIdx>
Moves the mask segment vertically. The specified offset is added to the y-values of all
points of the selected mask segment.
Only takes effect after the MTESt:SEGMent:RESCale:RECalculate command.
Parameters:
<OffsetY> Range: -1000 to 1000
Increment: 1E-6
*RST: 0
Default unit: V
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
<MaskSegmIdx> Number of the segment. Counting starts from 0.
Usage: Asynchronous command
22.14.3 Mask definition: waveform mask
The chapter contains commands required for the definition of waveform masks -
MTESt:CTYPe is set to WFML.
MTESt:REFWfm.......................................................................................................... 1469
MTESt:WFMLupdate....................................................................................................1470
MTESt:WFMRescale:XWIDth........................................................................................1470
MTESt:WFMRescale:YWIDth........................................................................................1470
MTESt:WFMRescale:YPOSition.................................................................................... 1471
MTESt:WFMRescale:YSTRetch.................................................................................... 1471
MTESt:REFWfm <MaskTestName>,<Source>
MTESt:REFWfm? <MaskTestName>
Sets the reference waveform from which the mask is created.
The reference waveform can be created before, or loaded from a file with REFCurve
commands, see Chapter 22.10.2, "Reference waveforms", on page 1327.
Parameters:
<Source> REF1 | REFerence1 | REF2 | REFerence2 | REF3 |
REFerence3 | REF4 | REFerence4
REF1 = REFerence1, REF2 = REFerence2, REF3 = REFer-
ence3, REF4 = REFerence4: reference waveforms
*RST: REF1
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
Usage: Asynchronous command
Mask testing