Remote control commands
R&S
®
RTP
1250User Manual 1337.9952.02 ─ 12
Usage: Asynchronous command
22.8.13 High definition mode
HDEFinition:STATe.......................................................................................................1250
HDEFinition:BWIDth.....................................................................................................1250
HDEFinition:RESolution?.............................................................................................. 1250
FORMat:BORDer.........................................................................................................1251
HDEFinition:STATe <State>
Activates the high definition mode of the instrument.
Parameters:
<State> ON | OFF
ON: high definition mode, up to 16 bit digital resolution
OFF: normal oscilloscope mode
*RST: OFF
Usage: Asynchronous command
HDEFinition:BWIDth <Bandwidth>
Sets the filter bandwidth for the high definition mode.
Parameters:
<Bandwidth> Range: 10000 to max. 2 GHz, depending on the instru-
ment bandwidth.
Increment: 1000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
See "Bandwidth" on page 156 for bandwidth limits.
HDEFinition:RESolution?
Returns the resulting digital resolution in high definition mode.
Return values:
<Resolution> Range: 0 to 16
Increment: 0.1
*RST: 0
Default unit: bit
Usage: Query only
Asynchronous command
Acquisition and setup