Remote control commands
R&S
®
RTP
1204User Manual 1337.9952.02 ─ 12
Example:
ACQuire:COUNt 1
SINGle
CHANnel2:OVERload?
Queries the overload status of channel 2 after a single acquisi-
tion.
CHANnel2:OVERload OFF
Resets the overload status bit.
Usage: Asynchronous command
22.8.6 Waveform data
To set the export data format, see FORMat[:DATA] on page 1154.
CHANnel<m>[:WAVeform<n>]:DATA:HEADer?............................................................... 1204
CHANnel<m>[:WAVeform<n>]:DATA[:VALues]?.............................................................. 1205
CHANnel<m>[:WAVeform<n>]:DATA:HEADer?
Returns the header of channel waveform data.
Table 22-1: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of one waveform 200000
4 Number of values per sample interval. For most
waveforms the result is 1, for peak detect and enve-
lope waveforms it is 2. If the number is 2, the num-
ber of returned values is twice the number of sam-
ples (record length).
1
If multichannel export is enabled, the number of returned samples is Record length *
Number of exported waveforms. See also EXPort:WAVeform:MULTichannel.
Suffix:
<m>
.
1..4
Selects the input channel.
<n> 1..3
Selects the waveform. If [:WAVeform<n>] is omitted, waveform 1
is adressed.
Example:
CHAN1:WAV1:DATA:HEAD?
-9.477E-008,9.477E-008,200000,1
Usage: Query only
Acquisition and setup