EasyManua.ls Logo

Rohde & Schwarz RTP Series - Acquisition and Setup

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1186User Manual 1337.9952.02 ─ 12
Return values:
<DeviceFootprint> Content of the device footprint xml file as block data
DIAGnostic:SERVice:FWVersion?
Returns the firmware version that is currently installed on the instrument.
Return values:
<FirmwareVersion> Version string
Usage: Query only
DIAGnostic:SERVice:PARTnumber
<MaterialNumber>
Returns the material number of your instrument. This number is required to order a
new option, and in case of service.
Parameters:
<MaterialNumber> Number string
DIAGnostic:SERVice:SERialnumber?
Returns the serial number of your instrument. This number is required to order a new
option, and in case of service.
Return values:
<SerialNumber> Number string
Usage: Query only
DIAGnostic:SERVice:CHANnelcount?
Queries the number of available channels.
Return values:
<ChannelCount> Range: 0 to 4
Increment: 1
*RST: 0
Usage: Query only
Firmware/software: V 2.00
22.8 Acquisition and setup
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
Acquisition and setup

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals