EasyManua.ls Logo

Rohde & Schwarz RTP Series - Aux out

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1249User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..4
The suffix is irrelevant. The setting affects all active channels.
Parameters:
<ExtendSkew> ON | OFF
*RST: OFF
Usage: Asynchronous command
22.8.12 AUX OUT
CALibration:SOURce:FREQuency................................................................................. 1249
CALibration:SOURce:STATe..........................................................................................1249
CALibration:SOURce:FREQuency <Frequency>
Frequency value and waveform type of the internal calibration source, which is output
at the [Aux Out] connector.
Parameters:
<Frequency> KHZ7900 | MHZ10 | KHZ15800 | MHZ31_25 | KHZ31600 |
KHZ63200 | GHZ1 | MHZ250MB | MHZ250mb | GHZ1_20DB |
MHZ62 | MHZ125 | MHZ250 | MHZ500 | KHZ505500 | MHZ1R |
MHZ31R | MHZ31_25R | MHZ498S | MHZ1257_2S |
MHZ2491S | MHZ4060S | MHZ4350S | MHZ4810S |
MHZ4991S | MHZ5940S | MHZ1R | MHZ31R | MHZ31_25R |
MHZ498S | MHZ1257_2S | MHZ2491S | MHZ4350S |
MHZ4810S | MHZ4991S | MHZ4060S | MHZ5940S
MHZ1R | MHZ31R | MHZ31_25R | MHZ498S | MHZ1257_2S |
MHZ2491S | MHZ4350S | MHZ4810S | MHZ4991S | MHZ4060S
| MHZ5940S
Square wave 1MHz, 31 MHz, 31.25 MHz
Sine wave 498 MHz, 1257.2 MHz, 2491 MHz, 4350 MHz, 4810
MHz, 4991 MHz, 4060 MHz, 5940 MHz
*RST:
MHZ1R
Example:
CAL:SOUR:FREQ GHZ1
CAL:SOUR:STAT ON
The commands activate the 1 GHz reference frequency.
Usage: Asynchronous command
CALibration:SOURce:STATe <State>
Defines the state of the internal calibration source.
Parameters:
<State> ON | OFF
*RST: OFF
Acquisition and setup

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Related product manuals