This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
The following figure shows a cursor measurement on a spectrum waveform:
If "Peak excursion" is 30 dB, the peaks 1 to 5 are found. If "Peak excursion" is 20 dB,
also the peaks 6 to 10 are found. The cursor position is on peak 6.
Remote command:
CURSor<m>:PEXCursion on page 1507
MEASurement<m>:SPECtrum:PEXCursion on page 1394
8.2 Automatic measurements
The R&S RTP can perform many measurements in parallel. The measurements are
combined in up to 8 measurement groups. In addition, you can run the quick measure-
ment.
The basic measurement settings are source, category, and the selection of the mea-
surement. You can refine the setup to get more specific results:
Gating
A gate limits the measurement to a user-defined part of the waveform. See Chap-
ter 8.2.3, "Measurement gates", on page 339.
Statistics and long term measurements
To evaluate time-dependent behavior of measurement results, you can use statis-
tics, long term measurements, and tracks. You can also decide, how many mea-
Automatic measurements