Loading...
Measurements
R&S
®
RTP
329User Manual 1337.9952.02 ─ 12
Remote command:
CURSor<m>:SOURce on page 1350
Absolute
Both cursors are set to the absolute peak value.
Remote command:
CURSor<m>:MAXimum[:PEAK] on page 1358
Next absolute
Cursor 2 is set to the next smaller absolute peak from the current position.
Remote command:
CURSor<m>:MAXimum:NEXT on page 1359
Next left
Cursor 2 is set to the next peak to the left of the current position.
Remote command:
CURSor<m>:MAXimum:LEFT on page 1359
Next right
Cursor 2 is set to the next peak to the right of the current position.
Remote command:
CURSor<m>:MAXimum:RIGHt on page 1359
Center frequency
Sets the vertical cursor line Cu1 to the center frequency.
Remote command:
CURSor<m>:FFT:TOCenter on page 1358
Set center frequency to Cursor
Sets the center frequency to the frequency value that is measured at cursor line Cu1.
Remote command:
CURSor<m>:FFT:SETCenter on page 1358
Threshold
Defines an absolute threshold as an additional condition for the peak search. Only
peaks that exceed the threshold are detected.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
Remote command:
CURSor<m>:THReshold on page 1508
MEASurement<m>:SPECtrum:ATHReshold on page 1394
Peak excursion
Defines a relative threshold, the minimum level value by which the waveform must rise
or fall to be considered as a peak. To avoid identifying noise peaks, enter a peak
excursion value that is higher than the noise levels.