Mask testing
R&S
®
RTP
438User Manual 1337.9952.02 ─ 12
The general mask test settings are described in Chapter 10.2, "Mask test settings",
on page 426.
10.3.2 Eye mask definition
Access: [App Cockpit] > "Analysis" tab > "Define Mask" > "Mask Type" = "Eye" >
"Setup mask"
The "Eye Mask " dialog provides all settings to define masks of different shapes. It is
available if option R&S RTP-K12 is activated on the instrument.
Shape
Defines the outline of the eye mask: square, diamond, hexagon or octagon.
Remote command:
MTESt:EYEMask:TYPE on page 1476
Horizontal center/Vertical center
Set the horizontal (time) and vertical (voltage) values of the eye shape enter and thus
define the position of the eye shape on the display.
Remote command:
MTESt:EYEMask:HPOSition on page 1479
MTESt:EYEMask:VPOSition on page 1479
Mask testing on eye diagrams