Triggers
R&S
®
RTP
223User Manual 1337.9952.02 ─ 12
Remote command:
TRIGger<m>:WINDow:RANGe on page 1264
Upper level
Sets the upper voltage limit for the window.
Remote command:
TRIGger<m>:LEVel<n>:WINDow:UPPer on page 1264
Lower level
Sets the lower voltage limit for the window.
Remote command:
TRIGger<m>:LEVel<n>:WINDow:LOWer on page 1264
Time condition
Selects how the time limit of the window is defined. Time conditioning is available for
the vertical conditions "Stay within" and "Stay outside".
"Within"
Triggers if the signal stays inside or outside the vertical window limits
at least for the time Width - Delta and for Width + Delta at the most.
"Outside"
"Outside" is the opposite definition of "Within". The instrument triggers
if the signal stays inside or outside the vertical window limits for a
time shorter than Width - Delta or longer than Width + Delta.
"Shorter"
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
"Longer"
Triggers if the signal crosses vertical limits after the specified "Width"
time is reached.
Remote command:
TRIGger<m>:WINDow:TIME on page 1265
Width
For the ranges "Within" and "Outside", the width defines the center of a time range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time
lapse, respectively.
Remote command:
TRIGger<m>:WINDow:WIDTh on page 1265
±Delta
Defines a range around the "Width" value.
Remote command:
TRIGger<m>:WINDow:DELTa on page 1266
6.3.7 Timeout
Access: "Menu">"Trigger" > "Setup" tab > "Type = Timeout"
Trigger types