Triggers
R&S
®
RTP
243User Manual 1337.9952.02 ─ 12
Trigger out signal setup
Available, if "Trigger out" is set to "On trigger".
Defines the pulse that is provided to the [Trigger Out] connector on the rear panel.
A trigger out pulse can be provided either when a trigger occurs, or when a mask test
violation occurs, or when a limit check violation in a measurement occurs.
"Polarity"
Sets the polarity of the trigger out pulse, that is the direction of the
first pulse edge.
"Delay"
Sets the delay of the first pulse edge to the trigger point. The setting
is only available if "Enable trigger out" is active.
"Pulse length"
Sets the length of the trigger out pulse.
Remote command:
TRIGger<m>:OUT:POLarity on page 1313
TRIGger<m>:OUT:PLENgth on page 1314
TRIGger<m>:OUT:DELay on page 1314
6.7 Qualification
Access: "Menu" > "Trigger" > "Qualify" tab
Qualification

Table of Contents