Remote control commands
R&S
®
RTP
1307User Manual 1337.9952.02 ─ 12
See also:
●
TRIGger<m>:HOLDoff:MODE
●
TRIGger<m>:HOLDoff:AUTotime? on page 1306
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<AutoTimeScaling> Range: 1E-3 to 1000
Increment: 1
*RST: 0.5
Usage: Asynchronous command
22.9.18
Conditioning
TRIGger<m>:LEVel<n>:NOISe[:STATe]..........................................................................1307
TRIGger<m>:LEVel<n>:NOISe:MODE........................................................................... 1308
TRIGger<m>:LEVel<n>:NOISe:ABSolute....................................................................... 1308
TRIGger<m>:LEVel<n>:NOISe:PERDivision...................................................................1309
TRIGger<m>:LEVel<n>:NOISe:RELative........................................................................1309
TRIGger<m>:LEVel<n>:NOISe[:STATe] <HysteresisMode>
Selects how the hysteresis is set.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
<n> 1..31
Indicates the trigger source:
1...4 = channel 1 to 4
5 = external trigger input
6...9 and 11 = not available
10 = line trigger
12...27 = R&S RT-ZVC input channels
28...31 = differential signals Diff1, Diff2 and Common1, Com-
mon2
Parameters:
<HysteresisMode> AUTO | MANual
AUTO
This is the recommended mode. The hysteresis is set by the
instrument to reject at least the internal noise of the instrument.
You can define a higher minimum value using TRIGger<m>:
LEVel<n>:NOISe:ABSolute.
MANual
The hysteresis is defined directly with TRIGger<m>:
LEVel<n>:NOISe:ABSolute.
Trigger