Acquisition and waveform setup
R&S
®
RTP
198User Manual 1337.9952.02 ─ 12
5.7.2  ZVC voltage settings
Access: [App Cockpit]> "Analysis" > "Multi-channel probe"> "Setup" tab > "Setup" icon
of voltage channel
See also:
●
Scale
●
Offset
Channel
Selects the voltmeter channel to be configured.
State
Enables the corresponding voltage channel of the probe. The number of available
channels depend on the characteristics of your multi-channel power probe.
Remote command: 
ZVC:Z<m>:V<n>[:STATe] on page 1239
Position
Moves the selected signal up or down in the diagram. The visual effect is the same as
for Offset. While the offset sets a voltage, position is a graphical setting given in divi-
sions. Within a given operation range, modifying the position is equivalent to moving
vertically the display range. The position can only be modified such that the display
range reaches at most the limits of the operation range.
For the voltmeter channels, the Scale, the Offset and the "Position" specify the operat-
ing range of the voltmeter.
Remote command: 
ZVC:Z<m>:V<n>:POSition on page 1238
R&S
 RT-ZVC probe