Remote control commands
R&S
®
RTP
1180User Manual 1337.9952.02 ─ 12
22.7.3 Display
22.7.3.1 Persistence
DISPlay:PERSistence[:STATe].......................................................................................1180
DISPlay:PERSistence:INFinite.......................................................................................1180
DISPlay:PERSistence:TIME.......................................................................................... 1180
DISPlay:PERSistence:RESet.........................................................................................1180
DISPlay:PERSistence[:STATe] <State>
If enabled, each new data point in the diagram area remains on the screen for the
duration defined using DISPlay:PERSistence:TIME, or as long as DISPlay:
PERSistence:INFinite is enabled.
If disabled, the signal value is only displayed as long as it actually occurs.
Parameters:
<State> ON | OFF
*RST: ON
Usage: Asynchronous command
DISPlay:PERSistence:INFinite <State>
If persistence is enabled (DISPlay:PERSistence[:STATe]), each new data point in
the diagram area remains on the screen infinitely until this command is set to "OFF".
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
DISPlay:PERSistence:TIME <Time>
If persistence is enabled (DISPlay:PERSistence[:STATe]), each new data point in
the diagram area remains on the screen for the duration defined here.
Parameters:
<Time> Range: 0.05 to 50
Increment: 0.05
*RST: 0.05
Default unit: s
Usage: Asynchronous command
DISPlay:PERSistence:RESet
Resets the display, removing persistent values.
Instrument setup