Remote control commands
R&S
®
RTP
1190User Manual 1337.9952.02 ─ 12
TRIGger<m>:OFFSet:LIMited <State>
If ON, the horizontal position cannot be set outside the visible waveform diagram.
See also: TIMebase:HORizontal:POSition on page 1189
Suffix:
<m>
.
1..3
The numeric suffix is irrelevant.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
AUToscale
Performs an autoset process: analyzes the enabled channel signals, and obtains
appropriate horizontal, vertical, and trigger settings to display stable waveforms.
Rohde & Schwarz does not recommend using the autoset in remote control. To adjust
the oscilloscope remotely, especially for automated testing applications, use the
remote commands that adjust the horizontal, vertical and trigger settings.”
Usage:
Event
Asynchronous command
ACQuire:POINts:AUTO <ReclgthManual>
Selection to keep constant either the resolution or the record length when you adjust
the time scale (TIMebase:SCALe) or acquisition time (TIMebase:RANGe).
Parameters:
<ReclgthManual> RESolution | RECLength
RESolution
Resolution is kept constant. Set the required resolution value
with ACQuire:RESolution.
RECLength
The record length is kept constant. Set the required record
length value with ACQuire:POINts[:VALue].
*RST: RESolution
Usage: Asynchronous command
Acquisition and setup