Remote control commands
R&S
®
RTP
1233User Manual 1337.9952.02 ─ 12
Usage:  Asynchronous command
ZVC:Z<m>:I<n>:OVERload:RSTO <Settings>
Resets the overload indication at the probe.
In internal shunt mode (ZVC:Z<m>:I<n>:SHUNt:MODE is set to INTShunt), chooses
whether to keep the operation range adjusted by the R&S RT-ZVC during overload or
to restore the original settings for the given current channel.
In external shunt mode (ZVC:Z<m>:I<n>:SHUNt:MODE is set to EXTShunt), there is
no adjustment of the operation range during overload.
Suffix: 
<m>
.
1..2
<n> 1..4
Setting parameters: 
<Settings> ADJusted | ORIGinal
ADJusted
The operation range that is automatically adjusted by the probe
during overload is kept.
In internal shunt mode, during an overload at a current channel,
the probe switches automatically to the next higher range. If the
10A range is also exceeded, the amperemeter switches to exter-
nal shunt mode to protect the probe against permanent damage.
ORIGinal
The original operation range is restored as before the overload
and the adjustment of the operation range.
Usage: 
Setting only
Asynchronous command
ZVC:Z<m>:I<n>:OVERload:VALue?
Queries if an overload of the current channel was detected.
Suffix: 
<m>
.
1..2
<n> 1..4
Return values: 
<Overload> ON | OFF
*RST:  OFF
Usage:  Query only
Asynchronous command
ZVC:Z<m>:I<n>:POSition <VertPosi>
Sets the vertical position of the indicated current channel as a graphical value.
Acquisition and setup