Remote control commands
R&S
®
RTP
1305User Manual 1337.9952.02 ─ 12
Usage: Asynchronous command
TRIGger<m>:HOLDoff:EVENts <Events>
Defines the number of triggers to be skipped. The next trigger only occurs when this
number of events is reached. The setting is relevant if the holdoff mode is set to
EVENts.
See also:
TRIGger<m>:HOLDoff:MODE
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Events> Range: 1 to 2147483647
Increment: 1
*RST: 1
Example:
TRIGger:HOLDoff:MODE EVENts
TRIGger:HOLDoff:EVENts 5
Usage: Asynchronous command
TRIGger<m>:HOLDoff:MIN <RandomMinTime>
Defines the lower limit for the random time holdoff. The setting is relevant if the holdoff
mode is set to RANDom.
See also:
TRIGger<m>:HOLDoff:MODE
TRIGger<m>:HOLDoff:MAX
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<RandomMinTime> Range: 100E-9 to 5
Increment: 200E-6
*RST: 1E-3
Default unit: s
Example:
TRIGger:HOLDoff:MODE RANDom
TRIGger:HOLDoff:MIN 1ms
TRIGger:HOLDoff:MAX 2ms
The holdoff time is set randomly between 1 ms and 2 ms.
Usage: Asynchronous command
Trigger

Table of Contents

Other manuals for Rohde & Schwarz RTP Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTP Series and is the answer not in the manual?

Rohde & Schwarz RTP Series Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTP Series
CategoryTest Equipment
LanguageEnglish

Related product manuals