Remote control commands
R&S
®
RTP
1591User Manual 1337.9952.02 ─ 12
TRIGger<m>:SPI:DPTO............................................................................................... 1592
TRIGger<m>:SPI:FCONdition....................................................................................... 1592
TRIGger<m>:SPI:MISopattern.......................................................................................1593
TRIGger<m>:SPI:MOSipattern...................................................................................... 1593
TRIGger<m>:SPI:MODE <Type>
Selects the trigger type for SPI analysis.
Parameters:
<Type> SSACtive | TIMeout | MOSI | MISO | MOMI
SSACtive
Start of the message: slave select signal SS changes to the
active state.
TIMeout
Triggers on the next message start after the "Timeout" time.
MOSI
Triggers on a specified data pattern in that is expected on the
MOSI line. Define the pattern with TRIGger<m>:SPI:
MOSipattern.
MISO
Triggers on a specified data pattern in that is expected on the
MISO line. Define the pattern with TRIGger<m>:SPI:
MISopattern
MOMI
Triggers on a specified data patterns on the MISO and MISO
lines.
*RST:
SSACtive
Usage: Asynchronous command
TRIGger<m>:SPI:PALignment <DataAlignment>
Defines how the specified data pattern is searched.
Parameters:
<DataAlignment> WORD | BIT
WORD
The pattern is matched only at word boundaries.
BIT
Bit-by bit: the pattern can be at any position in the data word.
*RST: WORD
Usage: Asynchronous command